The BLENDER TM2 is a medical device designed for the controlled administration of a mixture of medical Air and Oxygen (FiO₂) to the patient.

Function: precise adjustment of the Oxygen concentration, from 21% to 100%, in medical Air.
Connection: supplied via two hosepipes (medical Air and Oxygen), connected either to wall-mounted pressurised gas sources or to quick-release outlets on regulators.

Features

  • O₂ concentration adjustment (FiO₂): from 21% to 100%

  • Accuracy: ± 3% of full scale

  • Connections: 2 NIST-compliant inlets (O₂ and medical Air), supplied via low-pressure hosepipes for direct connection to gas sources

  • Safety: integrated non-return system at the inlets, preventing any gas backflow into the network

  • Alarm: built-in audible alarm in case of pressure difference between O₂ and medical Air

    • Sound level: 80 dB at 30 cm

  • Reliability: 100% mechanical system, no electronics

  • Mounting options:

    • Rail-mounted

    • Wall-mounted

    • Vertical pole mounting with dedicated clamp

    • Mobile trolley

  • Traceability: each product engraved with a unique serial number

Versions

  • 3 available versions: Low Flow, High/Low Flow, High Flow